IN VIVO ENTRANCE DOSE MEASURMENTS IN HEAD AND NECK CANCER TREATMENT USING THERMOLUMINESCENT DOSIMETERS

Authors

  • Abobaker El-Toomi Radiotherapy and Nucl. Medicine Dep., National cancer Institute of Sabratha Author
  • Abir El-Bousefi Radiotherapy and Nucl. Medicine Dep., National cancer Institute of Sabratha Author

DOI:

https://doi.org/10.66411/jer.v25i.325

Abstract

For performing  in vivo  entrance  dose  measurements  in  external photon beam radiotherapy, characteristics of thermoluminescent dosimeters (LiF: Mg; Ti), known as Harshow TLD-100, studies were carried out. The TLD system (reader and dosimeters) stability and reproducibility were investigated and the calibration of the TLD's  was performed. The reference TLD was calibrated under a reference condition (100 cGy with Co-60;  SSD  80cm;   10×10  cm2),  using  30×30  cm2   Perspex  slices.  ATOM®   female phantom studies for head and neck were performed. A dose of 100 cGy was delivered to the isocenter at the center of one sit of the phantom  slices. The dose values “dose distribution” at different points were obtained. Entrance dose for 25 patients with 35 treatments for head and neck cancer were performed. The results were compared with the expected Dose values, only three treatments with discrepancy greater than 5% were observed.

201 87

References

[1] Claudio, F., Handbook of Thermoluminescence., World Scientific Publishing Co Ptl, Ltd, 2003

[2] Van dam, J. and Marinello G., Methods For In Vivo Dosimetry in External Radiotherapy, Brussels, (Belgium), 2006.

[3] TECHNICAL REPORTS SERIES NO.398 IAEA Viena, 2000

[4] ICRU, 2007

[5] Claudio C. et al, Thermoluminescent chip detector for in vivo dosimetry in pelvisand head & neck cancer treatment, INAC, Sautos, Brazil, 2007.

[6] Procedures for In Vivo Dosimetry with Solid State Detectors: IAEA publication, 200.

[7] PERE. ENGSTRÖMandet. al, Invivo dose verification of IMRT treated head and neck cancer patients, Acta Oncologica, 2005. DOI: https://doi.org/10.1080/02841860500218983

[8] User Manual ATOM® female phantom.

[9] User Manual TLD reader (RA’94).

Downloads

Published

30-03-2018

Issue

Section

Articles

How to Cite

[1]
A. El-Toomi and A. El-Bousefi, “IN VIVO ENTRANCE DOSE MEASURMENTS IN HEAD AND NECK CANCER TREATMENT USING THERMOLUMINESCENT DOSIMETERS”, JER, vol. 25, pp. 19–28, Mar. 2018, doi: 10.66411/jer.v25i.325.